Performance Specifications
External AC/DC 5V requirements (connection to USB)
Output voltage
Power output
Isolation
Classification
External AC/DC 48V requirements (connection to RJ45)
Output Voltage
Power output
Compliancy
Isolation
Classification
Connector
IntelliVue CL Hotspot Specifications
Physical Specifications
Size (W X H X D)
Weight
Robustness
Environmental Specifications
Operating Temperature Range
Operating Humidity Range
Operating Altitude Range
Storage / Transportation Temperature
Range
Storage / Transportation Humidity
Range
Storage / Transportation Altitude Range -500 to 4600m
5V +/-5%
min. 8 W
double according to IEC60950-1 Class II
Limited Power Source according to IEC60950-1
48V +/-10%
min. 8 W
IEEE 802.3af (PoE Class 3)
double according to IEC60950-1 Class II
Limited Power Source according to IEC60950-1
RJ45 (Data-in), RJ45 (Data&Power Out)
75 mm x 53 mm x 177 mm ±5%
3.0 in x 2.1 in x 7.0 in ±5%
330 g ±10%
11.6 oz ±10%
Provides essential performance during exposure to random vibration
according to IEC TR 60721-4-7 Class 7M1
Survives shock and 0.05 m free fall according to IEC TR 60721-4-7 Class
7M1
0 to 40°C (32 to 104°F)
≤95 %RH @ 40°C (104°F)
no condensation
-500 to 3000m
-500 to 2000 m when powered from external power supply 865222 #E90
-20 to 60°C (-4 to 140°F)
≤90% RH @ 60°C (140°F)
(non-condensing)
10 Specifications