Integrated Battery Handling
8
Battery Care
Battery care begins when you receive a new IntelliVue Cableless Measurement Device with built-in
battery and continues throughout the life of the device. The table below lists battery care activities and
when they should be performed.
Activity
Perform a visual inspection
Charge the battery
Store the device in a state of charge in
the range of 40-50% and power off the
device completely.
Handling Precautions
Lithium ion batteries store a large amount of energy in a small package. Use caution when handling the
IntelliVue Cableless Measurement Devices with built-in battery; misuse or abuse could cause bodily
injury and/or property damage.
•
Handle with care.
•
Do not expose the device to liquids.
•
Do not attempt to disassemble the device.
•
Do not put device in autoclave.
•
Damaged devices should not be used anymore.
WARNING
•
Do not crush or puncture - mechanical abuse can lead to internal damage and internal short
circuits which may not be visible externally.
•
Do not incinerate the devices or expose them to temperatures above 60°C (140°F).
When to perform
Before using the IntelliVue Cableless Measurement Device
Upon receipt, after use, or if a low battery state is indicated.
To optimize performance, a fully (or almost fully)
discharged battery should be charged as soon as possible.
When not in use for an extended period of time.
8