INOP Message, Indication
No Cradle
Numeric is replaced by -?-
INOP tone
No Host Monitoring
INOP tone
No System
Remove From Pat
INOP tone
Service Battery
INOP tone
<SpO₂ Label> Equip Malf
Numeric is replaced by -?-
INOP tone
<SpO₂ Label> Erratic
Numeric is replaced by -?-
INOP tone
<SpO₂ Label> Extd.Update
Numeric is replaced by -?-
<SpO₂ Label>Interference
Numeric is replaced by -?-
INOP tone
<SpO₂ Label> Low Perf
Numeric is replaced by -?-
<SpO₂ Label> NeoPatient?
Numeric is replaced by -?-
INOP tone
<SpO₂ Label> No Pulse
Numeric is replaced by -?-
INOP tone
Source
What to do
NBP
The NBP Pod is not in the cradle. You can silence this
INOP, but the INOP message remains visible until the
NBP Pod is inserted into the cradle and the next
measurement is started or the
selected.
If INOP appears with the NBP Pod inserted into the
cradle, remove and replace the cradle.
Cableless
There is a problem with the communication to the
Measurement Device
assigned patient monitor and monitoring is currently not
possible (no patient alarms or information). Check the
connection. Contact your service personnel.
Cableless
There is a problem with the communication to the
Measurement Device
network and central monitoring is currently not possible.
Check the connection.
Cableless
Displayed on the Cableless Measurement Device. The
Measurement Device
temperature of the battery is too high. Remove the
Cableless Measurement Device from the patient and
contact service personnel.
Battery
Maximum number of charge/discharge cycles for battery
exceeded. Charging of the battery is inhibited. Contact
your service personnel to replace battery.
SpO
The SpO
2
SpO
Check the sensor placement. Try another adapter cable
2
and sensor. If the INOP persists, contact your service
personnel.
SpO
The update time for displayed values is extended due to
2
an NBP measurement on the same limb or an excessively
noisy signal.
SpO
There is too much interference, caused by a high level of
2
ambient light and/or electrical interference. Cover the
sensor to minimize ambient light. If the INOP persists,
make sure that the sensor cable is not damaged or
positioned too close to power cables.
SpO
Accuracy may be compromised due to very low
2
perfusion. Stimulate circulation at sensor site. If INOP
persists, change the measurement site.
SpO
The patient monitor the SpO
2
neonatal mode. The INOP will remain active until the
monitor is changed to adult or pediatric mode.
SpO
Check the perfusion at the measurement site. If
2
necessary, stimulate circulation or change the
measurement site. If the INOP is due to an NBP
measurement on the same limb, wait until the NBP
measurement is finished.
5 Technical Alarms (INOPs)
Stop All
Pod is faulty. Contact your service personnel.
2
Pod is assigned to is in
2
SmartKey is
51