5 Technical Alarms (INOPs)
INOP Message, Indication
<SpO₂ Label> No Sensor
Numeric is replaced by -?-
INOP tone
<SpO₂ Label> NoisySignal
Numeric is replaced by -?-
INOP tone
<SpO₂ Label> Poor Signal
<SpO₂ Label> Pulse?
Numeric is replaced by -?-
INOP tone
<SpO₂ Label> Searching
Numeric is unavailable
<SpO₂ Label> Sensor Malf
Numeric is replaced by -?-
INOP tone
<SpO₂ Label> Sensor Off
Numeric is replaced by -?-
INOP tone
<SpO₂ Label> Unkn.Sensor
Numeric is replaced by -?-
INOP tone
<SpO₂ Label> Upgrade
Numeric is replaced by -?-
52
Source
What to do
SpO
Make sure the SpO
2
persists, try another sensor. If you acknowledge this
INOP, the measurement will be switched off.
SpO
Excessive patient movement or electrical interference is
2
causing irregular pulse patterns. Try to reduce patient
movement or to relieve the cable strain on the sensor.
SpO
The signal quality of the SpO
2
accuracy may be compromised.
SpO
The detectable pulsations of the SpO
2
the specified pulse rate range.
SpO
The patient signal is analyzed, but no valid numerics are
2
available yet.
SpO
The SpO
2
INOP persists, contact your service personnel.
SpO
The algorithm has determined that a sensor is connected,
2
but not properly applied to the patient. The ability of the
algorithm to detect this condition depends on the used
sensor type.
SpO
The connected SpO
2
measurement hardware.
SpO
SpO
2
sensor is connected. If the INOP
2
2
sensor is faulty. Try another sensor. If the
2
sensor is not supported by this
2
in upgrade mode, no patient monitoring possible.
2
measurement is poor. The
signal are outside
2