Displaying A List Of Current Inops; Setting The Volume Of The Inop Tone - Philips IntelliVue CL SpO2 Instructions For Use Manual

Table of Contents

Advertisement

5 Technical Alarms (INOPs)
This will silence the INOP tone and clear the INOP message. If the condition which caused the INOP
is still present, the INOP symbol will be displayed on the right hand side of the screen.
When using a Cableless Measurement Pod with an IntelliVue Information Center iX Release A, make
sure to remove the pod at the telemetry device and only when in SRR range of the telemetry device.
Otherwise you will get a technical alarm (
silenced at the telemetry device. You will then have to reboot the telemetry device to remove the
message.

Displaying a List of Current INOPs

To display a list of the currently active INOPs,
Select the
1
Select
2
3
Select

Setting the Volume of the INOP Tone

To set the volume for the INOP tone,
1
Select the
Select
2
Select
3
your configuration.
NOTE
No INOP tone will sound at the device as long as it is within the SRR range.
48
Main Setup
SmartKey.
INOPs
.
INOP Messages
.
Main Setup
SmartKey.
INOPs
.
INOP Vol
and select a volume setting. The maximum is 10 and the minimum depends on
If the volume of the INOP tone is set to zero, the following
symbol is displayed on the right icon tray of the Cableless
Measurement Device screen:
Full length INOP message
1
INOP indicator
2
cl NBP Disconnect
cl SpO₂ Disconnect
/
) that can not be

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the IntelliVue CL SpO2 and is the answer not in the manual?

Subscribe to Our Youtube Channel

This manual is also suitable for:

Intellivue cl npb

Table of Contents

Save PDF