Reference List of all INOPs

INOP Message, Indication
Batt Incompatible
INOP tone
Batt Malfunction
INOP tone
Battery Empty
INOP tone
Battery Low
INOP tone
Check Batt Temp
INOP tone
Check Battery
INOP tone
Check Charger I/F
INOP tone
Check Settings
INOP tone
cl NBP Disconnect
at Information Center or at patient
monitor
cl SpO₂ Disconnect
at Information Center or at patient
monitor
Source
What to do
Battery
Battery cannot be used with this Cableless Measurement
Device. Replace battery with one that has been approved
for use with this Cableless Measurement Device. Contact
your service personnel.
Battery
Malfunction of the battery system detected (charger
circuit or battery). Contact your service personnel.
Battery
The remaining monitoring time is below 30 minutes.
Charge battery.
Battery
The remaining monitoring time is below 2 hours.
Battery
The temperature of the battery is critically high. Check
that Cableless Measurement Device is not covered and
not exposed to a heat source. If INOP persists, remove
Cableless Measurement Device from patient and contact
your service personnel.
Battery
The maximum number of charge/discharge cycles of the
battery will be reached in less than 50 cycles. Contact
your service personnel to replace the battery.
Battery
Overvoltage or undervoltage detected at the charger
interface. Clean contacts of charger interface at Cableless
Measurement Device and charging station. If the INOP
persists, contact your service personnel.
Cableless
If this INOP appears and an INOP tone sounds, check
Measurement Device
the Cableless Measurement Device and patient settings
before you resume making measurements. If the settings
are unexpected, there may be a problem with the
Cableless Measurement Device software. Contact your
service personnel.
If this INOP is acknowledged at the Cableless
Measurement Device, it is cleared. If it is silenced
remotely, only the tone is cleared.
Cableless
The NBP Pod has lost the SRR connection to the
Measurement Device
telemetry device or the patient monitor.
Cableless
The SpO
Measurement Device
telemetry device or the patient monitor.
5 Technical Alarms (INOPs)