Philips IntelliVue CL SpO2 Instructions For Use Manual page 75

Table of Contents

Advertisement

IntelliVue CL NBP Pod Performance Specifications
Battery
Runtime (fully charged battery)
Battery Recharge Time
Short Range Radio Specifications
Type
Technology
Frequency band
Modulation
Bandwidth
Effective radiated power (ERP)
Electronic Article Surveillance (EAS) EAS tag inside the housing
Real Time Clock Accuracy
Measurement Ranges
Pressure Transducer Accuracy
(0 to 300 mmHg)
Blood Pressure Measurement
Accuracy
Pulse Rate Measurement Range
Pulse Rate Measurement Accuracy
Measurement Time
STAT Mode Cycle Time
Integrated rechargeable Li-Ion battery with battery gauge and cycle counter
Typical 4 days @ 2 measurements per hour
Typical 5 days @ 1 measurement every 2 hours
Maximum 2.5 hours
built in interface with integrated antenna
IEEE 802.15.4
2.4 GHz ISM (2.400 - 2.483 GHz)
DSSS (O-QPSK)
5 MHz
max. 0 dBm (1 mW)
less than 5 seconds per day (typically) as long as device is in power state
"Device on" or "Device off". Automatically synchronized with assigned
host system.
Adult:
Systolic: 30 to 270 mmHg (4.0 to 36.0 kPa)
Mean: 20 to 255 mmHg (2.5 to 34.0 kPa)
Diastolic: 10 to 245 mmHg (1.5 to 32.0 kPa)
Pediatric:
Systolic: 30 to 180 mmHg (4.0 to 24.0 kPa)
Mean: 20 to 160 mmHg (2.5 to 21.0 kPa)
Diastolic: 10 to 150 mmHg (1.5 to 20.0 kPa)
±3 mmHg @ 15 to 25 °C
±(3 mmHg or 2% whichever is greater) @ 10 to 40°C
According to ANSI/AAMI SP10:2002/(R)2008 + A1:2003
8 mmHg standard deviation
±5 mmHg mean error
40 to 300 bpm
40 - 100 bpm: ±5 bpm
101 - 200 bpm: ±5% of reading
201 - 300 bpm: ±10% of reading
(average over NBP measurement cycle)
Auto/manual/sequence mode:
Typical 40 seconds @ >60 bpm and normal adult cuff
Maximum 180 seconds
STAT Mode:
Typical 30 seconds @ >60 bpm and normal adult cuff
Maximum 180 seconds
5 minutes
10 Specifications