IntelliVue CL Transmitter Base Station
(shown here with the transmitter inserted)
The IntelliVue CL Transmitter Base Station is used with the IntelliVue CL Transmitter. When the
patient is in their room, the Transmitter can be inserted into the Base Station where it is powered and
recharged and also remains connected to the WLAN to communicate vital signs data.
WARNING
The IntelliVue CL Transmitter Base Station may not be used in the patient vicinity.
Ensure that the CL Transmitter Base Station does not come into close contact with implanted
pacemakers, to avoid magnetic interference affecting the mode of the pacemaker.
CAUTION
The CL Transmitter Base Station generates a magnetic field. Do not store magnetic media (such as
identity cards or credit cards with magnetic strips or magnetic tapes/disks) near to the Base Station, as
the data may be damaged.
IntelliVue CL Charging Station
The IntelliVue CL Charging Station is used to charge the IntelliVue CL SpO
NBP Pod and the IntelliVue CL Transmitter. The MX40 wearable patient monitor can also be charged
on the IntelliVue CL Charging Station. It contains nine charger slots for IntelliVue Cableless
Measurement Devices. Depending on its size a device may occupy one or two charger slots.
As soon as a device is placed onto a charger slot, the device will switch on and the charging cycle
begins automatically.
1 Introduction and Basic Operation
Pod, the IntelliVue CL
2
19