2 Monitoring or Collecting Vital Signs
approximately 20-25 seconds. Once a channel is identified, the first LED will light up and blink once
per second to indicate that the telemetry device is ready for assignment.
Add To
Mon 1
Mon 2
Tele 33
Tele 44
Select a patient monitor or telemetry system using the ◄ and ► keys.
1
If you select a patient monitor, the measurement selection key on that monitor will change to show
the type of measurement device.
Activate the assignment by pressing the key twice on the measurement device.
2
The cableless measurement device is assigned to the selected patient monitor or telemetry device.
A telemetry device plays the assignment tone when the assignment is successful. A patient monitor
issues an assignment prompt message.
If the internal measurement in the patient monitor is active (the measurement selection key is yellow),
you will need to confirm that it should be deactivated in favor of the cableless measurement device you
want to assign. To do this:
1
Select the measurement selection key on the monitor.
A prompt message appears with the
Select
2
When the cableless measurement device is assigned, the
that alarming capability has been transferred to the host (to the monitor or, for the telemetry device, to
the Information Center). No patient alarms will be announced on the Cableless Measurement Device.
If the device was assigned, this opens a menu for the patient monitor or telemetry system. To
unassign the measurement device from that monitor or telemetry system, select
confirmation the SRR connection is disconnected.
Assignment at the Patient Monitor
The cableless measurement device must be prepared for assignment by activating the
SmartKey.
At the patient monitor,
Open the
1
Cbleless
The available cableless measurement devices are shown in the window.
Select the device which you want to assign to the patient in the monitor.
2
The monitor displays the assignment prompt message.
3
If the internal measurement in the patient monitor is active, you will need to confirm that it should be
deactivated in favor of the cableless measurement device you want to assign.
When the cableless measurement device is assigned, the
that alarms from the device will be sent to the patient monitor.
An assigned cableless measurement device can be removed in the
For more details see the Instructions for Use for your patient monitor.
26
Confirm
to deactivate the internal measurement.
Add Cableless
window, e.g. using the measurement selection key followed by the
pop-up key.
Confirm
Cancel
and
keys.
symbol appears on its display indicating
symbol appears on its display indicating
Measurement Selection
Remove From
. After
Add/Remove
Add
window.
Need help?