Two-Stage Gas Pressure Regulators

NOTICE
NOTICE
Thermo Scientific
The quality of the CR gas regulator is critical to ensure optimum
performance of the instrument. A high quality two-stage pressure
regulator is absolutely essential, and it must be qualified for the purity
specification of the CR gas.
A gas regulator with two installed pressure gauges is not necessarily a
two-stage gas pressure regulator. Inform your local supplier that a
two-stage gas pressure regulator is absolutely essential and is a
requirement for stable performance of the iCAP Qnova Series
instruments.
The CR gas regulator should have a specification of 0.3 MPa (3 bar)
outlet pressure (max. 0.8 MPa [8 bar]), and it must fit the maximum
primary pressure specification of the CR gas laboratory supply. Check
whether your maximum primary pressure is 30 MPa (300 bar),
23 MPa (230 bar), or 20 MPa (200 bar), and choose accordingly.
Each country has a different standard outlet fitting for its gas supplies
and therefore each country will require a different regulator. Examples
for regulators are given by region in
listed, contact your local Thermo Fisher Scientific service office for
advice.
Table 5-22. Regulator suppliers
Country
Manufacturer
Europe
Linde
Spectrolab
US
Matheson
BOC
Tip Thermo Fisher Scientific does not endorse any manufacturer, nor
does it endorse products other than its own. Companies and products
listed in this guide are given as examples only.
iCAP Qnova Series ICP-MS Pre-Installation Requirements Guide (P/N BRE0006946, Revision A)
Gas Supply iCAP TQ/TQs ICP-MS
Table
5-10. If your region is not
Reference Code
Gas Application
C 202/2
Plasma
C 106/2
CR
C 200hv/2
CR
LM62
Plasma
3813-CGA580
CR
HP1700
CR
(CGA580)
Consumables