Pure Hydrogen or Mixtures with Helium (for QCell)
Oxygen or Mixtures with Helium (for QCell)
Thermo Scientific
Table 5-18. Gas requirements for argon (for QCell)
Gas Requirement
Supply Rate
Pressure
See
Table 5-19
for details of the hydrogen gas requirements.
Table 5-19. Gas requirements for hydrogen (for QCell)
Gas Requirement
Purity
Maximum Water Content
Supply Rate
Pressure
See
Table 5-20
for details of the oxygen gas requirements. All mixture
ratios are permitted.
Table 5-20. Gas requirements for oxygen (for QCell)
Gas Requirement
Purity
Maximum Water Content
Supply Rate
Pressure
iCAP Qnova Series ICP-MS Pre-Installation Requirements Guide (P/N BRE0006946, Revision A)
Gas Supply iCAP TQ/TQs ICP-MS
Specification
max. 12 mL/min
min. 0.05 MPa (0.5 bar)
max. 0.15 MPa (1.5 bar)
using a two-stage pressure
regulator (200 bar to 3 bar)
Specification
99.999%
2 ppmv
max. 15 mL/min
min. 0.05 MPa (0.5 bar)
max. 0.15 MPa (1.5 bar)
using a two-stage pressure
regulator (200 bar to 3 bar)
Specification
99.999% or better
2 ppmv
max. 2 mL/min for pure gas
min. 0.05 MPa (0.5 bar)
max. 0.15 MPa (1.5 bar)
using a two-stage pressure
regulator (200 bar to 3 bar)
Consumables
5-21