Operating Environment
Exhaust System
NOTICE
3-4
iCAP Qnova Series ICP-MS Pre-Installation Requirements Guide (P/N BRE0006946, Revision B)
Figure 3-2. Exhaust ports joined with Y-piece (option)
A purge connection is provided in the Y-piece of the exhaust line, see
Figure
3-2. This connection can be used when purging the CR gas
capillaries, see for example
Figure 3-3. Y-piece with capillary at purge connection
Table 3-2.
Specifications for the exhaust
Exhaust type
Hose ID [mm]
Plasma
63
Instrument
63
heat
Joined
100
a
exhaust
a
optional
The exhaust system must be tested for leakage before you connect it to
the instrument.
purge connection
Figure
3-3.
Exhaust Velocity [m/s]
6 to 8
4 to 6
5 to 7
3
Exhaust Flow [m
/h]
67 to 90
45 to 67
150 to 200
Thermo Scientific