Consumables
Permitted Materials
Table 5-28. iCAP RQ ICP-MS: Gases for CCT use
Material
iCAP RQ ICP-MS 2-Channel Systems
Hydrogen or mixtures with helium
Ammonia mixtures with helium
a
If the torque plate of your system does not mention "suitable for Hydrogen use," the specifications listed for 1-channel systems apply to your system.
Table 5-29. iCAP TQ/TQs ICP-MS: Gases for CR use
Material
Helium
Oxygen or mixtures with helium
Hydrogen or mixtures with helium
Ammonia or mixtures with helium
NOTICE
5-28
iCAP Qnova Series ICP-MS Pre-Installation Requirements Guide (P/N BRE0006946, Revision A)
Used for
a
:
Used for the removal of polyatomic and isobaric interferences
by charge-transfer reactions and mass-shift reactions
(hydrogen adducts).
NOTICE
A hydrogen sensor must be installed in the vicinity of the
instrument.
Used for the removal of polyatomic and isobaric interferences
by mass-shift reactions (ammonia cluster adducts).
NOTICE
Maximum content of ammonia: 1%. If an ammonia sensor is
installed in the lab, 35%.
Table 5-29
ICP-MS and iCAP TQs ICP-MS instruments.
Used for
Used for the removal of polyatomic interferences by kinetic
energy discrimination.
Used for the removal of polyatomic and isobaric interferences
by mass-shift reactions (oxidation).
Used for the removal of polyatomic and isobaric interferences
by charge-transfer reactions and mass-shift reactions (hydrogen
adducts).
NOTICE
A hydrogen sensor must be installed in the vicinity of the
instrument.
Used for the removal of polyatomic and isobaric interferences
by mass-shift reactions (ammonia cluster adducts).
NOTICE
An ammonia sensor must be installed in the vicinity of the
instrument.
Additional CRC gases (methane, ethane, propane, fluoromethane,
tetrafluoromethane, nitrous oxide, carbon monoxide, carbon dioxide,
acetylene, nitrogen, neon, xenon, krypton, sulfur hexafluoride) are
also supported for the iCAP TQ/TQs ICP-MS. Prior to use, your local
Thermo Fisher Scientific representative must be contacted to support
you with all necessary information and to guarantee an optimal
installation and configuration of the instrument.