Site Preparation
Dimensions and Weights
2-8
iCAP Qnova Series ICP-MS Pre-Installation Requirements Guide (P/N BRE0006946, Revision B)
To dismount the lifting device
1. Unscrew the two clamps (3 in
2. Dismount the lifting device.
3. Repeat for the lifting device on the other side of the instrument.
4. Store away the clamps, washers (2 in
Figure
2-4) along with the lifting devices for later use.
To store away the handles
1. At the left hand side of the instrument, push down the spring (1 in
Figure
2-7) of the handle 2 in
and push the handle into its housing (3 in
2
Labeled Components: 1=spring, 2=handle, 3=handles in housing
Figure 2-7. Shoving handle into housing
To mount the drip tray
1. If required, at the right side of the instrument, pull the handles
outside.
2. To mount the drip tray (1 in
two bolts at the bottom of the drip tray into the notches of the
handles. See 2 in
Figure
Figure
2-4) of the lifting device.
Figure
Figure 2-7
by using an screw driver
Figure
1
3
Figure
2-8) onto the handles, place the
2-8.
2-4) and screws (1 in
2-7).