Thermo Scientific iCAP Qnova Series Pre-Installation Requirements Manual page 29

Table of Contents

Advertisement

Thermo Scientific
Figure 2-12
shows typical data system hardware components.
communication
cables USB and
Ethernet
Figure 2-12. Typical data system workbench
Table 2-2
lists space requirements and weights of the typical data system
hardware components. The actual values depend upon your equipment.
Table 2-2.
Typical data system space and load requirements
Module
Height [cm]
Monitor
36
Minitower
48
computer
The USB communication cables between the computer and the mass
spectrometer must be no longer than 3 m each. The workbench that
holds the data system must be located next to the workbench that holds
the mass spectrometer. For a connection length of more than 3 m, an
active hub must be placed between the instrument and the computer.
The length of the USB cable between the iCAP Qnova Series
instrument and the hub is limited to 3 m, whereas between the hub and
the computer a cable length of 5 m can be used. This hub must be used
exclusively for the control of the mass spectrometer.
iCAP Qnova Series ICP-MS Pre-Installation Requirements Guide (P/N BRE0006946, Revision B)
monitor
wall outlet for data system components
120 or 230 V AC
Width [cm]
41
18
Site Preparation
Dimensions and Weights
computer
Length [cm]
Weight [kg]
18
6
43
14