Technical Data iCAP TQ/TQs ICP-MS
Parameter
Instrument Dimensions
iCAP TQ/TQs ICP-MS
Power Requirements
iCAP TQ/TQs ICP-MS
Fore vacuum pump iCAP TQ
Sogevac™ SV40Bi
Fore vacuum pump iCAP TQs
ECODRY 65 plus
Operating Environment
Temperature
Exhaust
Heat generation
Humidity
Cooling Water Requirements
Cooling water
Plasma and Cooling Gas
Argon
Thermo Scientific
The table summarizes the most important technical data of the iCAP
TQ/TQs ICP-MS systems. See the respective chapters of the manual
and
"Permitted Materials"
instrument properties.
Specification
Length × width × height
Weight
Nominal Voltage
Power
Wire
Fuse
Protection type (acc. IEC 60529)
Voltage
Power
Voltage
Power
Laboratory temperature
Optimum operation temperature
Exhaust velocity (plasma)
Exhaust flow (plasma)
Exhaust velocity (heat)
Exhaust flow (heat)
Instrument
Fore vacuum pump
20-80%, non-condensing and non-corrosive atmosphere
Supply rate
Temperature
Pressure
pH
Recommended conductivity
Solid residual
Purity
Max. water content
Supply rate
Pressure
iCAP Qnova Series ICP-MS Pre-Installation Requirements Guide (P/N BRE0006946, Revision B)
on
page 5-27
for details and additional
Value
772 mm × 665 mm × 1102 mm
151 kg
200-240 V AC (50/60 Hz), single phase
apparent power: 3000 VA;
effective power: 2200 W
Single-phase, 2-pole, 3-wire grounding
15/16 A (Europe)
20 A (North America, Japan)
IP 20
200-240 V (50 Hz/60 Hz)
1.6 kW (50 Hz)
1.9 kW (60 Hz)
200-240 V (50 Hz/60 Hz)
1.2 kW
15 to 35 °C
18 to 21 °C
6 to 8 m/s
67 to 90 m³/h
4 to 6 m/s
45 to 67 m³/h
350 W-520 W (50 Hz)
350 W-520 W (60 Hz)
900 W (50 Hz)
1100 W (60 Hz)
> 5.5 L/min
20 to 30 °C, optimum at 21 °C
0.25 to 0.6 MPa (2.5 to 6 bar)
6-8
< 1000 μS/cm
< 50 μm particle size
99.996% or better
5 ppmv
max. 24 L/min
min. 0.55 MPa (5.5 bar)/max. 0.6 MPa (6 bar)