Thermo Scientific iCAP Qnova Series Pre-Installation Requirements Manual page 10

Table of Contents

Advertisement

Plasma and Cooling Gas
Argon
CCT Gas
Helium (for QCell)
Hydrogen or mixtures with helium
a
(for QCell)
Oxygen or mixtures with helium
(for QCell)
Ammonia or mixtures with helium
(for QCell)
Additional Gas
Argon
Pure helium or mixtures with
argon
Pure oxygen or mixtures with
helium and/or argon
Pure nitrogen or mixtures with
helium and/or argon
Methane mixtures with argon
a
Also see
"Gas Supply iCAP RQ ICP-MS"
b
If an ammonia sensor is installed in the lab, the max. content of ammonia is 35%.
viii
iCAP Qnova Series ICP-MS Pre-Installation Requirements Guide (P/N BRE0006946, Revision B)
Purity
Max. water content
Supply rate
Pressure
Purity
Max. water content
Supply rate
Pressure
Purity
Max. water content
Supply rate
Pressure
Purity
Max. water content
Supply rate
Pressure
Max. content of ammonia
Purity
Max. water content
Supply rate
Pressure
Purity
Max. water content
Supply rate
Pressure
Purity
Max. water content
Supply rate
Pressure
Purity
Max. water content
Supply rate
Pressure
Purity
Max. water content
Supply rate
Pressure
Max. content of methane
Purity
Max. water content
Supply rate
Pressure
on
page 5-5
and
"Permitted Materials"
99.996% or better
5 ppmv
max. 24 L/min
min. 0.55 MPa (5.5 bar)/max. 0.6 MPa (6 bar)
99.999% or better
2 ppmv
max. 10 mL/min per channel
min. 0.05 MPa (0.5 bar)/max. 0.15 MPa (1.5 bar)
99.999% or better
2 ppmv
max. 10 mL/min per channel
min. 0.05 MPa (0.5 bar)/max. 0.15 MPa (1.5 bar)
99.999% or better
2 ppmv
max. 10 mL/min per channel
min. 0.05 MPa (0.5 bar)/max. 0.15 MPa (1.5 bar)
b
1%
99.999% or better
2 ppmv
max. 10 mL/min per channel
min. 0.05 MPa (0.5 bar)/max. 0.15 MPa (1.5 bar)
99.996% or better
5 ppmv
max. 1 L/min per channel
max. 0.1 MPa (1 bar)
99.996% or better
5 ppmv
max. 1 L/min per channel
max. 0.1 MPa (1 bar)
99.996% or better
5 ppmv
max. 1 L/min per channel
max. 0.1 MPa (1 bar)
99.996% or better
5 ppmv
max. 1 L/min per channel
max. 0.1 MPa (1 bar)
4.5%
99.996% or better
5 ppmv
max. 1 L/min per channel
max. 0.1 MPa (1 bar)
on
page 5-27
for more information.
Thermo Scientific