Technical Data iCAP RQ 2-Channel Instrument
Parameter
Instrument Dimensions
iCAP RQ ICP-MS
Power Requirements
iCAP RQ ICP-MS
Fore vacuum pump
Operating Environment
Temperature
Exhaust
Heat generation
Humidity
Cooling Water Requirements
Cooling water
Thermo Scientific
The table summarizes the most important technical data of the
iCAP RQ 2-channel systems. See the respective chapters of the manual
for details and additional instrument properties.
Tip If the torque plate of your instrument (see
"suitable for Hydrogen use," the following information is applicable
for your system.
Specification
Length × width × height
Weight
Nominal Voltage
Power
Wire
Fuse
Protection type (acc. IEC 60529))
Voltage
Power
Laboratory temperature
Optimum operation temperature
Exhaust velocity (plasma)
Exhaust flow (plasma)
Exhaust velocity (heat)
Exhaust flow (heat)
Instrument
Fore vacuum pump
20-80%, non-condensing and non-corrosive atmosphere
Supply rate
Temperature
Pressure
pH
Recommended conductivity
Solid residual
iCAP Qnova Series ICP-MS Pre-Installation Requirements Guide (P/N BRE0006946, Revision B)
Value
772 mm × 665 mm × 1102 mm
142 kg
200-240 V AC (50/60 Hz), single phase
apparent power: 3000 VA;
effective power: 2200 W
Single-phase, 2-pole, 3-wire grounding
15/16 A (Europe)
20 A (North America, Japan)
IP 20
200-240 V (50 Hz/60 Hz)
1.6 kW (50 Hz)
1.9 kW (60 Hz)
15 to 35 °C
18 to 21 °C
6 to 8 m/s
67 to 90 m³/h
4 to 6 m/s
45 to 67m³/h
350 W-520 W (50 Hz)
350 W-520 W (60 Hz)
900 W (50 Hz)
1100 W (60 Hz)
> 5.5 L/min
20 to 30 °C, optimum at 21 °C
0.25 to 0.6 MPa (2.5 to 6 bar)
6-8
< 1000 μS/cm
< 50 μm particle size
Figure
5-5) states
vii