6.1.4
Messages in the status line
Various status messages can be displayed in the status line of the SIPROCESS UV600, as
listed in the following table:
Table 6- 2
Messages in the status line
Message
Description
Warm-up
The analyzer has not yet reached
its operating temperature
Measuring
The analyzer module is in the
normal operating state
Purging
A purging time is active, measuring
mode is interrupted
Cal/Val [...]
A calibration procedure or
validation measurement is
currently running
[current phase]
See logbook
A status or function message has
occurred
SIPROCESS UV600
Operating Instructions, 10/2013, A5E03529375-02
6.1 Control panel, display and control elements
Possible measures
⇒ Wait until "Measuring" is displayed.
☹ The measured values of the analyzer could not
yet be used.
☹ Do not carry out calibrations
☺ Other operating functions can be carried out.
Only appears if the menu branch of an analyzer module is
called.
If a calibration procedure or manual validation is being carried
⇒
out:
Continue the procedure in accordance with the instructions (see
Auto-Hotspot).
⇒ If a fully automatic calibration or validation is being carried out:
Wait until the automatic procedure is finished.
⇒ If the reason is unknown: Check the logbook (see )