Important functions (selection)
7.3 Calibration and validation functions
The procedure is finished when the message 'Display results?' appears. You can now:
● Leave the calibration procedure/validation measurement using the
● View the results using the
The following menu is displayed (example):
Meaning of the various parameters:
7.3.4

Display calibration results

You can use this function to display the result of the last calibration of the selected
component or auxiliary value
Menu call:
Main menu → Calibrations → Calibration → Measured component → Results → Zero/reference
point
112
Display on screen
Relative
Date
Time
Setpoint
Actual value
Absolute
key.
Description
Drift between this calibration and the previous calibration
Current date in representation YY-MM-DD
Current time in 24-hour representation
Setpoint of the used calibration gas
Mean value of the measured actual values
Absolute drift
Operating Instructions, 10/2013, A5E03529375-02
key
SIPROCESS UV600