Siemens SIPROCESS UV600 Manual page 135

Table of Contents

Advertisement

Range selection
The following output ranges are possible for the "Range selection" parameter:
0
1
auto
Autoranging responds as follows:
● If the current measured value is in the small output range (in the output range with the
● As soon as the measured value exceeds the full-scale value of the small output range,
● As soon as the actual measured value in the large output range falls below 90% of the
Digital output for status of output range
By selecting the "Status" parameter you access the following menu (example):
Use the
change the switching logic, press the
Set the desired switching logic (direct or inverted) there using the
keys.
SIPROCESS UV600
Operating Instructions, 10/2013, A5E03529375-02
Output range 0 permanently activated
Output range 1 permanently activated
Autoranging
smaller span), the small output range is automatically active.
the large output range is activated automatically (i.e. the electronic signal span of the
analog output is adapted).
full-scale value of the small output range, the small output range is activated again.
and
keys to select the desired digital output. If you subsequently wish to
Important functions (selection)
7.5 Parameter assignments
key to branch to the follow-up menu.
,
and
135