Important functions (selection)
7.5 Parameter assignments
Assigning the internal function (destination) to a digital input
In the previous menu, select the digital input. In the subsequent menu, select 'Destination'.
The following menu is displayed (example):
The parameters have the following meanings:
Name of calibration
Name of validation
Cancel Cal./Val.
Failure
Maintenance
Pump off
Calibration gas fault
Disable Cal./Val.
None
Prevents calibrations and validations
[1]
1. Select the desired destination using the
2. Search for the desired function which the digital input is to control in the follow-up menus.
148
Programmed automatic calibration
Programmed automatic validation measurement
Cancel a running calibration/validation
Activate status flag "F" (for complete system)
Activate status flag "M" (for complete system)
External control signal for switching off the gas pump
External fault message "Calibration gas not available"
External control signal to prevent calibrations/validations
Not occupied (electronic status of digital input has no effect)
,
and
keys.
Operating Instructions, 10/2013, A5E03529375-02
[1]
[1]
SIPROCESS UV600