Siemens SIPROCESS UV600 Manual page 170

Table of Contents

Advertisement

Technical data
10.1 Technical data
Electrical characteristics
Table 10- 3
Electrical characteristics
Power supply
(optional, see rating plate)
Line frequency (AC)
Permissible overvoltages (transients in the power
supply network)
Power consumption (analyzer)
Electrical safety
EMC– interference immunity
(electromagnetic compatibility)
Internal line fuse
- Primary
- Secondary
Measuring behavior
Table 10- 4
Measuring behavior
(relating to sample gas pressure 1013 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient
temperature)
Reference point drift
Zero drift
- Standard measuring ranges
- Small measuring ranges
(≤ 2x smallest measuring range)
- Measured components NO, NO
Repeatability
Linearity deviation
Detection limit (2σ)
- Standard measuring ranges
- Small measuring ranges
(≤ 2x smallest measuring range)
Influence of ambient temperature:
- Standard measuring ranges
- Small measuring ranges
(≤ 2x smallest measuring range)
Setting time (T
Starting time
170
, SO
2
2
)
90
93 ... 132 V AC
186 ... 264 V AC
47 ... 63 Hz
Up to overvoltage category II in accordance with
IEC 60364-4-443
Approx. 50 VA, max. 300 VA
Tested in accordance with EN 61010-1
In accordance with EN 61326-1, EN 61326-2-1,
EN 61000-6-2, EN 61000-6-4 and EU directive
2004/108/EU.
With irradiation of electromagnetic fields in the
frequency range 750 MHz ± 20 MHz, increased
errors in measurement may occur with small
measuring ranges
6.3 A, not replaceable
8 A
< ±1%/week of respective span
< ±1%/week of respective span
< ±2%/week of respective span
< ±1%/day of respective span
< ±1% of respective span
< ±1% of respective span
< 0.5% of span
< 1% of span
< 1%/10 K
< 2%/10 K
4 s
< 60 minutes
Operating Instructions, 10/2013, A5E03529375-02