Siemens SIPROCESS UV600 Manual page 99

Table of Contents

Advertisement

The following menu is displayed with the existing calibration gas settings (example).
In order to change the parameters of an existing calibration gas configuration, select them
using the
create a new calibration gas configuration, select a vacant location (#4 in the example) and
activate it by pressing the
The following menu is displayed with a calibration gas configuration (example):
Use the
the
The parameters have the following meanings:
Notation in the menu
Name
Active
Pump off
Valve
Use
SIPROCESS UV600
Operating Instructions, 10/2013, A5E03529375-02
and
keys and activate them by pressing the
and
keys to select the desired parameter, and activate it by pressing
key.
Function
Name of calibration gas configuration (or number of a vacant location)
Yes means that this calibration gas for calibrations/validations can be
selected
Status of sample gas pump during use of the calibration gas
Digital output which is activated during use of the calibration gas for
controlling a solenoid valve
Settings of calibration gas for measurement
7.3 Calibration and validation functions
key.
Important functions (selection)
key. In order to
99