The following menu is displayed with the existing calibration gas settings (example).
In order to change the parameters of an existing calibration gas configuration, select them
using the
create a new calibration gas configuration, select a vacant location (#4 in the example) and
activate it by pressing the
The following menu is displayed with a calibration gas configuration (example):
Use the
the
The parameters have the following meanings:
Notation in the menu
Name
Active
Pump off
Valve
Use
SIPROCESS UV600
Operating Instructions, 10/2013, A5E03529375-02
and
keys and activate them by pressing the
and
keys to select the desired parameter, and activate it by pressing
key.
Function
Name of calibration gas configuration (or number of a vacant location)
Yes means that this calibration gas for calibrations/validations can be
selected
Status of sample gas pump during use of the calibration gas
Digital output which is activated during use of the calibration gas for
controlling a solenoid valve
Settings of calibration gas for measurement
7.3 Calibration and validation functions
key.
Important functions (selection)
key. In order to
99