Important functions (selection)
7.5 Parameter assignments
Configure the "Span" parameter in the same manner.
7.5.4

Selecting output ranges

This function provides the following options:
● Determination of the analog output for output of a specific measured value
● Specification of a specific output range or both output ranges
● Specification of which digital output signals the current output range
These settings can be selected for each individual component (also for virtual components).
If measuring point switching is set, the settings can be made for each individual measuring
point.
Menu call:
Main menu → Parameters → Output ranges
The following menu is displayed (example):
Use the
by pressing the
The following menu is displayed (example):
You can now set the "Range selection" and "Status" parameters for the selected
components.
134
and
keys to select the desired measured component, and confirm this
key.
Operating Instructions, 10/2013, A5E03529375-02