11. Status

The electromagnetic flowmeter can detect and display various device or
application errors.
If there is a status or error message, the STATUS symbol in the display
alternately flashes orange / red. To call up the status / error information, the
status key must be pressed, then the status window that appears then lists all the
messages that have accumulated up to this point in time. By pressing the 
key, the user confirms the knowledge of the displayed errors, the status memory
is cleared and the status window is closed. If one of the displayed errors persists,
it will be reported again by flashing the status icon.
The following status / error messages are generated:
Display text
Empty Pipe
Temp Sens Error
Meas saturated
No Subslave
Simulation
page 38
Description
Measuring tube is not completely
filled with medium or medium
with too low conductivity is used.
Error in the temperature
measuring circuit
Flow measuring circuit
overdriven
Internal hardware error
Simulation function active
Debugging
Check the filling of the
measuring circuit or
conductivity of the medium
(> 20 μs / cm)
Siemens
Repair by
Service
necessary
Reduce flow rate
Siemens
Repair by
Service
necessary
-