Status Messages (Function 53) - Siemens FIDAMAT 6 Operating Instructions Manual

Gas analyzer for determination of total hydrocarbon content
Table of Contents

Advertisement

Functions
6.5 Parameters
6.5.5

Status messages (function 53)

Figure 6-28
With this function, you can configure the display of status messages in the status line of the
measured value display. You can display a maximum of three different status messages.
The message CODE ■/□ provides information on the current operating mode ("Coded
display mode"/"Decoded display mode") and is therefore always displayed.
The device activates the function control (softkey 5 in the function display) when it
determines that the measured value was influenced.
This is the case:
● During the warm-up phase
● After entry of the code (i.e. change to "Operator control mode")
● After remote control via the interface (ELAN, PROFIBUS)
● During automatic calibration (Autocal) or remote calibration
Note
Control function ("CTRL")
When the function control is active, the measured value can be invalid!
As long as you are in code level 1 or 2, thus in "Operator control mode", the function
control is active. If you have configured the analog output with "Store on" under
function 77, this remains in fault current state until you remove it because any code level
has to be exited.
86
Status messages
Operating Instructions, 03/2018, A5E00222135-03
FIDAMAT 6