Mounting and connecting
4.5 Electrical connections
Plug connector X7 – analog outputs (AO) and analog inputs (AI)
Analog outputs
Figure 4-8
Characteristic of analog outputs (see also Auto-Hotspot):
The analog outputs are floating (electrically isolated) and provide a load-independent current
signal.
● Signal range 0...24 mA
● Residual ripple 0.02 mA
● Resolution 0.1%
● Accuracy 0.25% of full-scale value
● Maximum load 500 Ω
● Maximum voltage ±50 V
● Adjustable start or fault state
Note
Electrical isolation
The electrical isolation is canceled if the negative poles of the analog outputs are
connected to GND
38
Plug connector X7 - analog outputs
Operating Instructions, 10/2013, A5E03529375-02