Technical data
10.1

Additional technical data

10.1.1
Input
The measured variable is the distance between the reference point (see Fig. on → ä 16) and the
product surface.
The level is calculated taking into account the empty distance entered (E, see Fig. on → ä 84).
Alternatively, the level can be converted to other variables (volume, mass) by means of linearization
(32 points).
10.1.2
Output
• 4 to 20 mA with HART protocol
Failure information can be accessed via the following interfaces:
• Local display:
– Error symbol (→ ä 37)
– Plain-text display
• Current output, failsafe mode can be selected (e.g. according to NAMUR Recommendation
NE 43).
• Digital interface
The Levelflex M linearization function enables the measured value to be converted into any desired
length or volume units and mass or %. Linearization tables for volume calculation in cylindrical
tanks are preprogrammed. Any other tables with up to 32 value pairs can be input manually or semi-
automatically. The creation of a linearization table with FieldCare is particularly convenient.
10.1.3
Performance characteristics
• Temperature = +20 °C ±5 °C
• Pressure = 1013 mbar abs. ±20 mbar
• Humidity = 65 % ±20%
• Reflection factor ≥ 0.8 (surface of the water for coax probe, metal plate for rod and rope probe
with min. 1 m ∅)
• Flange for rod or rope probe ≥ 30 cm ∅
• Distance to obstructions ≥ 1 m
• For interface measurement:
– Coax probe
– DK of the lower medium = 80 (water)
– DK of the upper medium = 2 (oil)
• Digital: 1 mm
• Analog: 0.03 % of the measuring range
Is in the Function group "Basic Setup" (00) as of → ä 48.
The reaction time depends on the configuration.
Shortest time:
• 2-wire electronics: 1 s
Levelflex M FMP40 with HART/4...20 mA