Ulvac UTM3400 Series Instruction Manual page 89

Turbo molecular pump
Table of Contents

Advertisement

Table 9-4 If the STATUS Orange Lamp Lights
Alarm
Alarm name
code
11
TD COUNTER LIMIT
12
PF COUNTER LIMIT
1
14
AC LOW VOLTAGE
15
POWER FAILURE
2
16
TMP:OVERLOAD
3
21
TMP TEMP/MB CABLE
24
TMP:OVER TEMP
4
22
TMP:SENSOR ERROR
46
MOTOR OVERSPEED
5
47
EI:R-SPEED ERROR
23
EI:MOTOR OVERCURR
34
EI:INV. OVERCURR
6
31
EI:BR OVERTEMP
32
EI:DC-DC OVERTEMP
7
35
EI:INV. OVERVOLT
36
EI:DC-DC LOW VOLT
8
37
EI:DC-DC OVERCURR
38
EI:DC-DC OVERVOLT
45
EI:BRAKE OVERTIME
9.5 Alarm Detection Capabilities
Possible cause
The number of high
The touch-down bearing may have
speed or power failure
deteriorated. Consult ULVAC or an
touch-downs has
approved service company regarding
exceeded the
replacement of the touch-down
prescribed number.
bearing.
Power failure or
Wait for the power to be restored. An
reduction in the power
unfamiliar sound will be heard a few
supply voltage.
minutes after a power failure. This
sound occurs when the rotor contacts
the protective bearing as magnetic
levitation cannot be maintained. This
is not an abnormal sound.
POWER switch was
Wait about 5 seconds before turning
turned off by mistake.
the POWER switch back on. Re-
acceleration is possible after resetting
and start-up.
Drop in rotation speed
Check that the outlet and inlet
during rotation at rated
pressures are below the specified
speed (increased
maximum pressures.
internal pump
Check for leakage.
pressure).
Check that too much purge gas is not
flowing.
Check that process gas flow rate is
not too high.
High pump unit
Check that the ambient temperature
temperature.
around the pump is within the
specified range.
Check that the temperature and flow
rate of the cooling water are within
the specified ranges.
Check that no load in excess of the
specified range is continuously
applied to the pump.
Rotation detection
Check that equipment causing noise
signal from the motor
is not used around the pump and
cannot be detected
power cable.
correctly.
Overcurrent supplied
Check that the ambient temperature
to the motor.
around the pump is within the
specified range.
Check that the temperature and flow
rate of the cooling water are within
the specified ranges.
Temperature increase
Check that the ambient temperature
in the control system.
around the pump is within the
specified range.
Check that the temperature and flow
rate of the cooling water are within
the specified ranges.
Defective circuit in the
Turn the power on again after the
power supply unit.
pump stops. The power supply unit
must be repaired if the problem
occurs again.
Sec-
Remedy
tion
4.1
4.1
5.2
4.1
5.3
4.1
5.3
9