Monitoring
Setting the upper alarm limit and the
pressure threshold
The standard alarm limits configured for the
ventilation mode can be used unchanged, see
chapter ''Changing the alarm limits'' on page 138.
or
The alarm limits can be set individually for the
current case:
1 Press the
(A) key.
Volume Control
The dialog window (B) with the alarm limits opens.
2 Adjust upper alarm limit and pressure threshold
of the peak pressure (PEAK) (C), see chapter
''Changing the alarm limits'' on page 138.
3 Confirm new values.
126
A
B
C
C
NOTE
The pressure threshold is preferably to be set so
that the value lies approx. 4 cmH
the current peak pressure.
Instructions for use Fabius GS premium SW 3.n
O (hPa) under
2