7 Reprocessing steps in accordance with ISO 17664

7.1 Overview of the reprocessing options

Manual
Manual dis-
cleaning
infection
Yes
Yes
Non-fixing
No
pre-cleaning
Non-fixing
No
pre-cleaning
Yes
Yes
Yes
Yes
Yes
Yes
No
No

7.2 Preparations at the site of use

Hazard from contaminated products.
Contaminated products are associated with an infection hazard.
▶ Take suitable personal protective measures.
▶ Pull off the USB plug of the medical device.
▶ Reprocess the medical device as soon as possible after treatment.
▶ The medical device must be dry when transported to reprocessing.
▶ To minimise the risk of infection during reprocessing, always wear protect-
ive gloves.
▶ Remove gross soiling immediately.
▶ Do not place in solutions or similar substances.

7.3 Disassembly

Incomplete reprocessing.
Infection hazard.
▶ To ensure complete reprocessing of all parts, camera, tip and Fluorescence
Cover need to be removed before reprocessing.
Instructions for use DIAGNOcam Vision Full HD
Automated
cleaning
No
Yes
Yes
No
No
No
No
WARNING
WARNING
37 / 52
Automated
Sterilisation
disinfection
No
No
Yes
Yes
Yes
Yes
No
No
No
No
No
No
No
No