7 Reprocessing steps in accordance with ISO 17664 | 7.1 Preparations at the site of use

7 Reprocessing steps in accordance with ISO 17664

7.1 Preparations at the site of use

Hazard from non-sterile products.
There is a risk of infection from contaminated medical devices.
▶ Take suitable personal protective measures.
▶ Remove all residual cement, composite or blood immediately.
▶ Reprocess the medical device as soon as possible after treatment.
▶ Remove the dental bur or diamond grinder from the medical device.
▶ The medical device must be dry when transported to reprocessing.
▶ Do not place in solutions or similar substance.

7.2 Cleaning

Never reprocess this medical device in an ultrasonic device.
Malfunction and material damage.
▶ Clean manually or in a washer disinfector only.

7.2.1 Manual external cleaning

Accessories required:
o
▪ Tap water 30
C ± 5
▪ Brush, e.g. medium-hard toothbrush
▶ Brush off under flowing tap water.

7.2.2 Automated external cleaning

KaVo recommends washer disinfectors in compliance with EN ISO 15883-1 that are
operated with alkaline cleaning agents.
The validations were conducted with the VARIO-TD program, the cleaning agent
neodisher® MediClean and the neutralisation agent neodisher® Z.
▶ The program settings and cleansers and disinfectants that must be used can be
found in the instructions for use of the thermodisinfector.
▶ Directly after automated cleaning/disinfection, treat the medical device with the
KaVo care products and systems provided by KaVo.

7.2.3 Manual internal cleaning

Validated internal cleaning (removal of residual protein) can be accomplished with
KaVo CLEANspray and KaVo DRYspray.