M3A Vital Signs Monitor User Manual
Technical alarms:
Message
TEMP EXCEED
LIMIT
The infrared ear thermometer will also give error messages on its screen. For details about the
error messages, refer to the accompanying operating instructions of the thermometer.
NOTE:
If the infrared ear thermometer frequently signals ERR alarms, the insulated board inside
the thermometer housing is malfunctioning or the ambient temperature changes, and the
monitor will delete the measurement values onscreen to avoid misoperation.

10.2.5 Replacing the Battery

The device is supplied with one lithium cell CR2032x1.
To replacing the battery, follow the procedure:
1.
Open the battery cover by inserting a pointed object into the battery cover pin hole;
meanwhile, use thumb to push battery cover out.
2.
Hold the thermometer and flip the battery out with a small screwdriver.
Insert the new battery under the metal hook on the left side ① and press the right side ②
3.
of the battery down until the it clicks in place.
Cause
The TEMP value is
beyond the range of
34 C --42.2 C.
Alarm Level
Check the integrity of the
probe cover, make sure it is
Medium
clean,
measurement.
- 73 -
TEMP Monitoring (Optional)
What to do
and
take
a
new