M3A Vital Signs Monitor User Manual
SpO
Setup
2
Sensitivity
The SpO
reading is the average of data collected within a specific time. You can set Sensitivity
2
to Low, Medium or High via the menu. The higher the sensitivity is, the quicker the monitor
responds to the changes in the patient's oxygen saturation level. Contrarily, the lower the
sensitivity is, the slower the monitor responds to the changes in the patient's oxygen saturation
level, but the measurement accuracy will be improved. When a critical patient is monitored,
selecting high sensitivity will help to understand the patient's state.
Sensor Off
You can configure the alarm level for SpO
Alarm Setup
Audio Alarm Paused
To activate the auditory alarm function, you can set Audio Alarm Paused to any value: 60 s,
120 s or 180 s, then press the button
information area displays the remaining pause time in seconds. Key sounds and beat sounds
keep their previous status. Auditory alarm will resume when you press the button
again or the pause time terminates.
Audio Alarm Off
To activate the Audio Alarm Off function, you can set Audio Alarm Off to ON, then press
the button
auditory alarm. Key sounds and beat sounds keep their previous status. Press the button
again to resume auditory alarm. If you set Audio Alarm Off to OFF, the function
is inactivated.
 Alarm Record (only available for the monitor outfitted with a recorder)
By configuring Alarm Record, the function of automatically outputting the alarm information in
case of any physiological alarm can be enabled or disabled. If the item is ON, the monitor will
automatically print out the alarm information once any physiological alarm is triggered. If the
item is OFF, the monitor will not automatically output the alarm information.
Other Setup
 Net Setup
Access Net Setup and you can see the menu:
Bed No.: Set the bedside monitor number to a value from 1~64.
Server IP
The default server IP is 202.114.4.119. It can be changed by the user according to the IP of PC
installed with MFM-CMS of EDAN.
Sensor Off to Low or High.
2
on the front panel and hold it for more than three seconds to turn off
- 31 -
on the front panel to pause audio alarm. The