M3A Vital Signs Monitor User Manual
2
The device must stay in stable ambient (room) temperature for 30 minutes before
operation.
3
Before the measurement, please stay in a stable environment for five minutes and
avoid exercise or bath for 30 minutes.
4
It is recommended that you measure the same ear for three times. If the three
measurements are different, select the highiest temperature.
5
Remember to compare the measurement result to the regular temperature of the
patient.
6
There is no gender and age limitation for using infrared ear thermometer.
7
The data saved in the thermometer is the last measurement data before the
thermometer is powered off.
Clinical repeatability: 0.12 C (1~5 years old); 0.10 C (> 5 years old).
8

10.2.3 TEMP Setup

Click on the TEMP Setup in the Main Menu:
TEMP Unit: Set temperature unit to C or ℉.

10.2.4 Alarm

10.2.4.1 Adjustable Range of Alarm Limits
The alarm limits are as follows:
Patient Type
Measure position
ADU/PED
Ear
10.2.4.2 Alarm Information and Prompts
Physiological alarms:
Message
TEMP HIGH
TEMP LOW
+35.5 C (+95.9 ℉)~+42 C (+107.6 ℉)
Measuring value of TEMP is above upper alarm limit.
Measuring value of TEMP is below lower alarm limit.
Alarm Limits
Cause
- 72 -
TEMP Monitoring (Optional)
Step
+0.1 C
Alarm Level
Medium
Medium