item to save the setup and return to the previous menu. If you want to exit the menu without
saving it, press the MENU button on front panel.

4.8.4 Default Configuration

NOTE:
Select any item in this submenu to cancel the current setup and use the selected default
setup. The one labeled with ■ is current configuration.
Factory Default Config: select the factory default configuration.
User Default Config: select the user-defined default configuration.
Restore Selected Config: select this item to restore the selected configuration.

4.9 Maintenance

Select Maintenance in Main Menu to open the menu. Factory Maintenance is only available
for the service personnel of EDAN or representatives authorized by EDAN.
User Maintenance
Input the user password 9 9 8 1 in the Enter Password box and press Confirm:
User Maintenance menu will pop up, in which you can set the following items.
Language: Set the displaying language.
NOTE:
The user should restart the monitor after changing the displaying language.
Nurse Call: Turn on or off the nurse call. When the parameter alarm occurs, the monitor
gives a 3s nurse call alarm prompt; if the audio alarm or the audio system is off, the monitor
can also give the nurse call alarm in abnormal condition.
The relay contact between pin7 and pin8 of RJ45 is normally open. But it is closed when an
alarm is audible.
NIBP Setup
NIBP Memory
You can set this item to ON or OFF. If the item is ON, the monitor will automatically memorize
the initial measurements of the patient when measuring his or her blood pressure. Then the
monitor will inflate the cuff according to the previous memorized measurements. This function
accelerates the measuring of the patient's blood pressure.
AVG Interval: Set the average NIBP measurement interval to 1min, 2min, 3min, 4min or 5min.
AVG Times: Set the average NIBP measurement times to 3 or 5.
NIBP Reset: select it to reset the NIBP module.
Restore measurement status.
Pick this item to restore initial settings of the pressure pump.
- 29 -