EDAN M3A User Manual page 39

Vital signs monitor
Hide thumbs Also See for M3A:
Table of Contents

Advertisement

M3A Vital Signs Monitor User Manual
Server Port: Set the server port.
 Barcode Setup
ID Start/ID End:
You can determine which digit in the barcode is the starting/ending digit for the patient ID via
configuration of ID Start and ID End.
Patient Type:
You can determine which digit in the barcode indicates the patient type. For example, if you set
Patient Type to 1, the first digit in the barcode will be identified as an indication of patient type.
Auto:
If it is set to ON, the patient information is updated automatically by using a barcode scanner. If
it is set to OFF, a message box indicating "Confirm to update patient, yes?"will pop up when
scanning a barcode. Click on YES to automatically update patient information; click on NO to
quit automatical update.
Adult/ Pediatry/ Neonate:
Select a digit from 0~9 to indicate the patient type.
NOTE:
1
Barcodes containing characters other than digits or containing space will be
considered invalid and cannot be identified. If any invalid character is detected, a
message box indicating "Special signs in barcode!" will appear.
2
Make sure the patient ID number obtained from the barcode is within the range of
patient ID stored in the monitor, or a message box indicating "Invalid patient ID!" will
appear.
3
Connect the barcode scanner to the monitor and wait 10 seconds before starting the
scanner.
4
Refer to the accompanying operator's manual of the scanner for more information
about its usage.
Save as User Default Config: save the current setup as the user default configuration;
Exit: Exit the menu.
Factory Maintenance
Factory maintenance function is only available for the service personnel of EDAN or
representative authorized by EDAN.
Calibrating NIBP
NIBP is not user-calibrated. NIBP pressure transducers must be verified at least once every two
years by a qualified service professional, and calibrated, if necessary. See the Service Manual for
details.
Version
Select Main Menu > Maintenance > About to check the version of the modules.