Temp Setup; Alarm - EDAN M3A User Manual

Vital signs monitor
Hide thumbs Also See for M3A:
Table of Contents

Advertisement

M3A Vital Signs Monitor User Manual
After Predict measuring is over, the measuring result displays, and MEASURE OVER appears
on the interface.
6 If the predict measuring is successfully finished, the monitor enters monitor mode after 30s;
otherwise the monitor enter monitor mode immediately after the predict measuring. The
monitoring state lasts for 10 min, and then the monitor enters waiting state.
TEMP parameter area on interface. Put the sensor back into the sensor bracket.
7 If necessary, repeat the measurement according to the procedure above.
NOTE:
1
After one measurement, the user should put the sensor back to the sensor bracket
and then take it out for starting a new measurement.
The monitor's state can change from the PREDICT mode into the MONITOR mode,
2
but it can not change from the MONITOR mode into the PREDICT mode.
3
The reference body site temperature is the same as the temperature of the
measuring site.

10.1.3 TEMP Setup

Click on the TEMP Setup in the Main Menu to display:
Monitor: when this item is selectable, select it to enter monitor mode.
Position: you can set this item to Oral, Axillary or Recta. The axillary sensor can be used
for measuring oral/axillary temperature, while the rectal sensor for measuring rectal
temperature.
TEMP Unit: Set temperature unit to C or ℉.

10.1.4 Alarm

10.1.4.1 Adjustable Range of Alarm Limits
The range for higher alarm limit and lower alarm limit is as follows:
Patient Type
Measure position
ADU
Oral/Axillary/Rectal
PED
Oral/Axillary/Rectal
Limit
+35.5 C (+95.9 ℉) ~ +42 C (+107.6 ℉)
+35.5 C (+95.9 ℉) ~ +42 C (+107.6 ℉)
- 67 -
TEMP Monitoring (Optional)
— — —
displays in the
Step
+ 0.1C