M3A Vital Signs Monitor User Manual
6 EXPLOSION HAZARD-Do not use the monitor in a flammable atmosphere where
anesthetics or other flammable materials may accumulate.
7 Do not attempt to connect or disconnect a power cord with wet hands. Make certain
that your hands are clean and dry before touching a power cord.
8 SHOCK HAZARD- the power receptacle must be a three-wire grounded outlet. A
hospital grade outlet is required. Never adapt the three-prong plug of the monitor to fit
a two-slot outlet.
9 Accessory equipment connected to the analog and digital interfaces must be certified
according to the respective IEC/EN standards. Furthermore all configurations shall
comply with the valid version of the system standard IEC/EN60601-1. Everybody who
connects additional equipment to the signal input connector or signal output
connector configures a medical system, and is therefore responsible that the system
complies with the requirements of the valid version of the system standard
IEC/EN60601-1. If in doubt, consult our technical service department or your local
distributor.
10 Use the battery only in this monitor. Do not connect battery directly to an electric outlet
or cigarette lighter charger.
11 Make sure the monitor is used in the appointed range of voltage, the effect of power
supply can be ignored.
12 Do not solder the leading wire and the battery terminal directly.
13 If liquid leaking from the battery gets into your eyes, onto your skin or clothes, do not
rub your eyes. Wash them well with clean water and go to see a doctor immediately.
14 Always keep the battery away from fire.
15 Stop using the battery if abnormal heat, odor, discoloration, deformation or abnormal
condition is detected during use, charge, or storage. Keep it away from the monitor.
16 Do not use a battery with serious scratch or deformation.
17 Only patient cable and other accessories supplied by EDAN can be used. Or else, the
performance and electric shock protection can not be guaranteed, and the patient
may be injured.
18 Please set the alarm according to the individual condition of patient to avoid delaying
treatment. Ensure there will be an alarm audio prompt when an alarm occurs.
19 Devices connecting with the monitor should be equipotential.
20 This monitor is not a device for treatment purposes.
21 The monitor is suitable for use in the presence of electrosurgery. When the monitor is
used with HF surgical equipment, user (doctor or nurse) should be cautious about
patient safety.
WARNING
- 3 -
Intended Use and Safety Guidance