Technical alarms:
Message
Temp
COMM
STOP
Temp
exceed
limit
No
TEMP
SENSOR
TEMP SENSOR
ERR
Temp Error P06
Temp Error E01
Temp Error E02
Temp Error E03
Temp Error E04
Heater error
Temp Error P02
Cause
TEMP
module
failure
or
communication
failure.
The TEMP value is
out of the range of
+30 C ~ +43 C.
Probe configuration
(or
no
probe
connected) error.
Incorrect sensor type
or incorrect measure
sites
Unable to pre-heat
probe tip.
System error during
synchronization.
System error during
patient
thermistor
calibration.
System error during
heater
thermistor
calibration.
System timing error.
Heater error.
Monitor Mode patient
thermistor unstable or
Alarm Level
Stop using measuring function
of
High
biomedical
Manufacturer's service staff.
Put the probe into the probe
Medium
well; take it out and measure
again.
Well connect the probe and the
Low
monitor, and measure again.
Use the correct type of sensor;
Medium
choose the correct measesure
site.
Stop using measuring function
of
biomedical
Manufacturer's service staff.
Low
NOTE: Measure readings
displayed on the screen are
unreliable when the monitor
indicates Temp Error P06.
Medium
Medium
Stop using measuring function
of
Medium
biomedical
Manufacturer's service staff.
Medium
Medium
Low
- 80 -
TEMP Monitoring (Optional)
What to do
TEMP
module;
engineer
TEMP
module;
engineer
TEMP
module;
engineer
notify
or
notify
or
notify
or