10.1.4.2 Alarm Information and Prompts
In monitoring mode, no physiological alarms are available.
Physiological alarms:
Message
TEMP HIGH
Measuring value of TEMP is above upper alarm limit.
TEMP LOW
Measuring value of TEMP is below lower alarm limit.
Technical alarms:
Message
TEMP
TEMP module failure or
COMM STOP
communication failure.
The
TEMP
beyond the range of +25
EXCCED
C ~ +45 C.
LIMIT
TEMP
TEMP NO
connected to the TEMP
SENSOR
module.
The Sensor temperature is
AMBIENT
higher than +40 C.
TEMP HIGH
The Sensor temperature is
AMBIENT
lower than +10 C.
TEMP LOW
The TEMP sensor cannot
TEMP
work properly.
SENSOR
ERR
Malfunction
TEMP
warm-up circuit.
HEATER
ERR
WARNING
Cause
Cause
TEMP
value
is
sensor
is
not
in
the
- 68 -
Alarm Level
Stop using measuring function
of
TEMP
High
biomedical
Manufacturer's service staff.
Put the sensor into the sensor
Medium
bracket,
measure again.
Connect the sensor and the
Low
monitor well, and measure
again.
Put the sensor into the sensor
bracket, measure again after
Low
the
reaches normal value.
Put the sensor into the sensor
bracket,
Medium
measure again. If the problem
persists, stop using measuring
function of TEMP module;
Medium
notify biomedical engineer or
Manufacturer's service staff.
TEMP Monitoring (Optional)
Alarm Level
Medium
Medium
What to do
module;
notify
engineer
take
it
out
ambient
temperature
take
it
out
or
and
and