M3A Vital Signs Monitor User Manual
Message
MEASURE
TIMEOUT
INIT
PRESSURE
TOO HIGH
PRESSURE
RANGE
EXCEED
Prompt message:
Message
Press NIBP START
Manual measuring...
Continual measuring...
Automatic measuring...
Measurement over
Calibrating...
Calibration over
Leakage testing...
Leakage test over
NIBP Resetting...
NIBP Alarm Off
Cause
Measuring
time
exceeded 120 s (adult)
or 90 s (neonatal).
The initial pressure is
too
high
during
measuring.
The measured pressure
exceeds the limit.
You can start NIBP measuring of continual mode.
During manual measuring mode.
During continual measuring mode.
During automatic measuring mode.
Measurement over
During calibrating
Calibration over
During leakage test
Leakage test over
NIBP module is resetting
The alarm of NIBP is turned off.
Alarm
Level
has
Measure again or use other
Low
measuring methods.
Measure again, if failure persists,
stop using measuring function of
Low
NIBP
biomedical
manufacturer's service staff.
Low
Measure by other method.
Cause
- 64 -
NIBP Monitoring (Optional)
What to do
module
and
notify
engineer