M3A Vital Signs Monitor User Manual
3
Only use the disposable probe covers supplied or recommended by EDAN. Use of
other manufacturer's probe covers, reuse of disposable probe covers or absence of
probe covers may produce temperature measurement errors and/or inaccuraies.
4
The infrared ear thermometer is not intended for neonatal patients.
This thermometer converts the ear temperature to display its 'oral equivalent'
5
(according to the result of the clinical evaluation to get the offset value). The
thermometer is adjusted to display an oral temperature equivalent. Oral Mode = Ear
Mode + 0.30 ° C.
1
Keep the probe window clean, dry, and undamaged at times to ensure accuate
measurements. To protect the probe window, always keep the thermometer in the
storage cover while transporting or when not in use.
2
Proper installation of the probe cover ensures accurate measurements.
3
Do not autoclave.
4
The thermometer is not waterproof. Do not immerse or drip fluids on it. Should this
occur, dry the thermometer with warm air. Check for proper operation and accuracy.
5
Holding the thermometer too long may cause a higher ambient temperature reading
of the probe, which could make the body temperature measurements lower than
usual.
6
Check whether the thermometer is damaged once it drops. If you cannot make sure
of it, send the complete device to your local dealer for recalibration.
7
Keep the unit dry and away from any liquids and direct sunlight.
8
For more details about using the infrared ear thermometer, refer to the
accompanying operating instructions of the thermometer.
9
The monitor outfitted with the TH module must not be used together with other
electrosurgery equipment, for example, ESU.
10 The probe should not be submerged into liquids.
11 Verification of the temperature module is necessary as frequently as dictated by your
Hospital Procedures Policy. When you need to calibrate the temperature
measurement, please contact the manufacturer.

10.2.2 Measuring Procedure

1.
Align the center of the probe to the center of the probe cover. Make sure to place the
adhensive side of probe cover upward.
2.
Insert the probe into the probe cover on the probe cover loader until the probe cover clicks in
place.
WARNING
CAUTION
- 70 -