EDAN M3A User Manual page 70

Vital signs monitor
Hide thumbs Also See for M3A:
Table of Contents

Advertisement

M3A Vital Signs Monitor User Manual
Technical alarms: (display in the area below the NIBP value):
Message
NIBP
COMM
STOP
NIBP INVALID
RESET
NIBP MODULE
ERR
LOOSE CUFF
AIR LEAK
AIR
PREESURE
ERR
NIBP SIGNAL
TOO WEAK
NIBP
NOISY
SIGNAL
OVER
PRESSURE
NIBP
SELFTEST
ERR
CUFF
TYPE
ERR
Cause
NIBP module failure or
communication failure.
The hardware pressure
is too high
The NIBP module has
failure.
Cuff is no properly
wrapped or no cuff
exists.
NIBP pump, valve, cuff
or tube has a leakage.
The airway of NIBP has
failure.
Cuff is too loose or
patient pulse is too
weak.
Because of arm motion,
signal noise is too large
or pulse rate is not
regular.
Pressure has exceeded
the
specified
upper
safety limit.
When the monitor is
powered
on,
NIBP
module is detected to
fail in calibration.
Cuff type does not
comply with the patient
type.
Alarm
Level
Stop using measuring function of
NIBP module; notify biomedical
High
engineer
service staff.
High
Measure again, if failure persists,
stop using measuring function of
NIBP
biomedical
High
manufacturer's service staff.
Low
Properly wrap the cuff.
Low
Check and replace the leaking
parts,
biomedical
manufacturer's service staff.
Low
Use other method to measure
Low
blood pressure.
Make sure that the patient under
Low
monitoring is motionless.
Measure again, if failure persists,
stop using measuring function of
Low
NIBP
biomedical
manufacturer's service staff.
High
Contact your service personnel.
Low
Select appropriate cuff type
- 63 -
NIBP Monitoring (Optional)
What to do
or
manufacturer's
module
and
notify
engineer
if
required,
notify
engineer
module
and
notify
engineer
or
or
or