Alarm Information And Prompts - EDAN M3A User Manual

Vital signs monitor
Hide thumbs Also See for M3A:
Table of Contents

Advertisement

M3A Vital Signs Monitor User Manual
DIA
10 mmHg ~180 mmHg
MAP
20 mmHg ~195 mmHg
Neonatal Mode
SYS
40 mmHg ~ 135 mmHg
DIA
10 mmHg ~ 100 mmHg
MAP
20 mmHg ~ 110 mmHg
Default NIBP alarm limits:
ADU (mmHg)
Lower
Limit
SYS
90
DIA
50
MAP
60
PR adjustable range of alarm limits:
PR
Default PR alarm limits:
Max. Upper Limit (bpm)
ADU
PED
NEO

9.6.2 Alarm Information and Prompts

Physiological alarms:
Message
NIBP SYS measuring value is above upper alarm limit.
SYS HIGH
NIBP SYS measuring value is below lower alarm limit.
SYS LOW
NIBP DIA measuring value is above upper alarm limit.
DIA HIGH
NIBP DIA measuring value is below lower alarm limit.
DIA LOW
MAP HIGH
NIBP MAP measuring value is above upper alarm limit.
MAP LOW
NIBP MAP measuring value is below lower alarm limit.
PED (mmHg)
Upper
Lower
Limit
Limit
160
70
90
40
110
50
Adjustable range
40~240
Min. Lower Limit (bpm)
120
160
200
Cause
Upper
Limit
120
70
90
50
75
100
- 62 -
NIBP Monitoring (Optional)
NEO (mmHg)
Lower
Upper
Limit
Limit
40
90
20
60
25
70
Step (bpm)
1
Step (bpm)
1
1
1
Alarm Level
Medium
Medium
Medium
Medium
Medium
Medium