10.3 TEMP Monitoring with F3000 Module

10.3.1 General Information

M3A with the F3000 module measures patient temperatures by oral, axillary or rectal means.
The electromagnetic compatibility of this device has been verified by test according to the
EN60601-1-2: 2015 requirements.
The monitor can only measure temperature of adult and pediatric patients. If the user measure
temperature of neonate patient, the monitor will not display data.

10.3.2 Safety Information

1
Do not use this device near flammable anesthetics. Not suitable for use in the
presence of flammable anesthetic mixture with air, oxygen or nitrous oxide.
2
Use probe covers supplied by the manufacturer with this thermometer only. Use of
any other probe cover will result in erroneous temperature readings.
3
Do not use this thermometer without first installing a new probe cover.
4
The thermometer and probe covers are Non-sterile. Do not use on abraded tissue.
5
To limit cross contamination, use Blue devices for Oral and Axillary temperature
taking only.
6
Use RED devices only for RECTAL temperatures.
7
Thoroughly dry all electrical contacts on both probe and thermometer after washing,
or device may fail to function properly.
8
For re-calibration, service or integrity checks, refer to a qualified Biomedical
Technician or return to the manufacturer.
9
Do not open the F3000 module. No user-serviceable parts inside. Opening of the
module may affect calibration and voids warranty.
10 Disposal of used probe covers must be performed in accordance with current
medical practices or local regulations regarding disposal of infectious, biological
medical waste.
11 Cleaning frequency and practices must be consistent with institutional policy for
cleaning of non-sterile devices.
12 The F3000 module is not intended for neonatal patients.
13 In monitoring mode, no physiological alarms are available.