M3A Vital Signs Monitor User Manual
 Other audible signals are not affected, including beat volume, key volume.
Pressing the button
When Audio Alarm Off is set to OFF, this function is inactivated. The monitor has no response
if you press the button
NOTE:
If a new alarm occurs during the audio alarm off period, the new alarm will not be
sounding.

5.5 Alarm Reset

To reset the alarm, you can select Main Menu > Alarm Setup > Alarm Reset. After resetting
the alarm,
 The audio alarm (including technical alarm and physiological alarm) is turned off, and no
alarms are sounding.
 The visual alarm indications are still displayed.
 It will not influence the configuration of physiological alarm off, audio alarm paused and
audio alarm off status.
NOTE:
If a new alarm occurs after the alarm is reset, the new alarm will be sounding.

5.6 Parameter Alarm

1
Prior to monitoring, make sure that the alarm limit settings are appropriate for your
patient.
2
Setting alarm limits to extreme values may cause the alarm system to become
ineffective.
In Main Menu > Alarm Setup, you can check and set the alarm limit or alarm status. The setup
is isolated from each other.
When a parameter alarm is OFF, an icon
turned off individually, they must be turned on individually.
For the parameters whose alarm is set to ON, the alarm will be triggered when at least one of
them exceeds alarm limits. The following actions take place:
1. Alarm message displays on the screen as described in alarm mode;
2. The monitor beeps in its corresponding alarm class and volume;
3. Alarm lamp flashes.
again can resume the audio alarm.
.
WARNING
- 38 -
displays near the parameter. If the alarms are
Alarm