52
7.7 Form of the rotation speed output signal
In the factory settings, two 90_ phase offset speed signals (5 V symmetrical,
complementary RS422 signals) are available. You can double the pulse count
set in each case by moving switch S6 to the "On" position. Pin 3 then outputs
the direction of rotation as a static direction of rotation signal (pin 3 = +5 V, pin
7 = 0 V opposite pin 8), if the shaft turns in the direction of the arrow. At a
rotation speed of 0 min
value.
Fig. 7.10: Direction of rotation arrow on the head of the sensor
Fig. 7.11: Switch for doubling the pulses
HBM
1
, the direction of rotation signal has the last measured
Switch S6
T10FS
Direction of
rotation arrow
Doubling the pulses
On
Off
S6
A0785 14.0 en