6.2 Calibration Option 2, code SF1/SU2

Applying a separated extra low voltage of 5 V to pin 6 (+) and 7 ( ) on plug 1
or 3 triggers the calibration signal.
The nominal (rated) voltage for triggering the calibration signal is 5 V (triggered
when U>2.7 V). The trigger voltage is electrically isolated from the supply
voltage and the measurement voltage. The maximum permissible voltage is 30
V
When voltages are less than 0.7 V, the torque flange is in measuring mode.
.
Current consumption at nominal (rated) voltage is approx. 2 mA and at
maximum voltage is approx. 22 mA.
NOTE
In the case of HBM system solutions, the measuring amplifier triggers the
calibration signal.